Brake Pin; Door Latch; Filter; Name Rating Plate - Beckman Coulter Airfuge Instructions For Use Manual

Air driven ultracentrifuge
Hide thumbs Also See for Airfuge:
Table of Contents

Advertisement

Description

Filter

Brake Pin

Braking action is provided by a movable brake pin (see
shaft of the stator (see
it 2 minutes after the timer reaches zero. When released, the pin engages the rotor bottom,
stabilizing the rotor as it gradually decelerates to a stop.

Door Latch

The door-latch mechanism secures the door during centrifugation. This prevents premature door
opening, which would result in damage to the rotor and stator pad.
Filter
A filter is supplied with the ultracentrifuge for water and oil removal from the air supply. The filter
element turns red when it is saturated and needs to be replaced.

Name Rating Plate

A name rating plate is affixed to the rear of the instrument. Always mention the serial number and
model number when contacting Beckman Coulter regarding your Airfuge ultracentrifuge.

Specifications

Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data,
without guarantee.
Table 1.1 Specifications
Specification
Speed range
Maximum relative centrifugal field
1-6
Figure
1.3). An electromagnet holds the pin down during a run and releases
Description
80 000 to 110 000 (5 000) rpm
199 000  g