7.3.3
High Breath Rate Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
94
Alarms
User manual
Description
High Breath Rate
Medium
A High Breath Rate alarm will be given when the alarm limit
has been exceeded for 15 seconds.
•
Mismatch between the Breath Rate setting and the alarm
setting.
•
Increased Breath Rate.
•
Too sensitive setting of the inspiratory trigger setting.
The breath rate goes below the alarm limit.
The Vivo 2 will continue treatment according to the current
settings.
•
10 bpm to 50 bpm
•
Off
1 bpm.
Doc. 007229 D-3