Breath Rate; Inspiratory Time (Insp. Time); Minimum Inspiratory Time (Min Insp. Time) - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

4.2.5

Breath Rate

The Breath Rate setting defines the minimum number of breaths the Vivo 3 will deliver
as long as no inspiratory trigger effort from the patient is detected. The cycles will be
ventilator-initiated breaths.
Property
Value range
Default value
Setting resolution
Available in modes
4.2.6

Inspiratory Time (Insp. Time)

The Inspiratory Time setting defines the length of each inspiration from start of inspira-
tion to cycling off to expiration.
Property
Value range
Default value
Setting resolution
Available in modes
4.2.7

Minimum Inspiratory Time (Min Insp. Time)

The Minimum Inspiratory Time setting defines a minimum length for each inspiration.
If the Minimum Inspiratory Time is set to Off, the length of the inspiration and/or
maximum inspiratory time is dependent on the set Expiratory Trigger.
Property
Value range
Default value
Doc. 007232 Y-2
Description
0 BPM to 40 BPM
The available range for Breath Rate might also be limited by
the set values for Inspiratory Time and Target Volume.
12 BPM
1 BPM
PCV+A
PSV
S
S/T
T
Description
0.3 s to 3 s
The available range for Inspiratory Time might also be
limited by the set values for Breath Rate.
1.5 s
0.1 s
PCV+A
PSV
S/T
T
Description
• Off
• 0.3 s to 3 s
Off
Treatment Functions and Settings 55
Vivo 3 Clinician's Manual

Advertisement

Table of Contents
loading

Table of Contents