5.4.2
Insp. Pressure
The Insp. Pressure setting is used to define the airway pressure during the inspiratory phase.
Minimum/maximum working IPAP is limited/achieved by a software control of blower
speed vs. measured pressure.
Min
Max
Default
Resolution
Tolerance
5.4.3
PEEP
The PEEP setting is used to define the airway pressure at the end of the expiratory phase.
Min
Max
Default
Resolution
Tolerance
5.4.4
Breath Rate
The Breath Rate setting defines the minimum number of breaths the ventilator will deliver
as long as no inspiratory trigger effort from the patient is detected. The cycles will be venti-
lator-initiated breaths.
The combination of the Breath Rate and Inspiratory Time setting is limited by the I:E ratio
range, 2:1 to 1:99.
Min
Max
Default
Resolution
Tolerance
5.4.5
Backup Rate
The Backup Rate setting defines the minimum number of breaths the ventilator will deliver
as long as no inspiratory trigger effort from the patient is detected. The cycles will be venti-
lator-initiated breaths.
Doc. 006150 M-1
O
4 cmH
2
60 cmH
O
2
15 cmH
O
2
0.5 cmH
O < 10 cmH
2
1.0 cmH
O ≥ 10 cmH
2
± (0.5 cmH
O + 5%)
2
2 cmH
O
2
Off (Off is only for dual limb circuits and active exhalation valve circuits)
30 cmH
O (Adult mode)
2
20 cmH
O (Paediatric mode)
2
For SIMV and pressure ventilation modes: The max setting is also
limited byInsp. Pressure -2 cmH
5 cmH
O (Adult mode)
2
2 cmH
O (Paediatric mode)
2
0.5 cmH
O < 10 cmH
2
1.0 cmH
O ≥ 10 cmH
2
± (0.5 cmH
O + 5%)
2
4 breaths/minute (Adult mode)
6 breaths/minute (Paediatric mode)
40 breaths/minute (Adult mode)
60 breaths/minute (Paediatric mode)
12 breaths/minute (Adult mode)
20 breaths/minute (Paediatric mode)
1
± 2%
O
2
O
2
O or SIMV Support Pressure - 2 cmH
2
O
2
O
2
O.
2
How to Use the Vivo 65
User manual
57