5.3.11

Total Rate

The Total Rate displays the actual total breath rate independent of whether the breaths are
patient- or ventilator-triggered breaths.
5.3.12

Spont Rate

The Spont Rate displays the actual spontaneous breath rate.
5.3.13

% Spont

% Spont displays the percentage of spontaneous breaths calculated since the ventilator was
last started (not calculated until 100 breaths are registered).
5.3.14
SpO
(Oxygen Saturation)
2
The SpO
accessory.
When using the ventilator with the SpO
tional oxygen saturation measured by the sensor.
The following information concerns the light emitted by the SpO
1. Peak Wavelength (red): 660 nm
2. Peak Wavelength (infrared): 905 nm
3. Maximum Optical Output Power: 15 mW
4. For more information regarding the oxygen probe's range of peak wavelengths, max opti-
cal power and usage, please refer to the respective probe manual.
Environmental factors may influence the function or accuracy of the pulse oximeter,
such as ambient light, physical movement, diagnostic testing, low perfusion, electromag-
netic interference, dysfunctional haemoglobin, presence of certain dyes and inappropri-
ate positioning of the pulse oximeter probe.
A functional test cannot be used to assess the Accuracy of a Pulse Oximeter Probe or a
Pulse Oximeter.
5.3.15

Pulse Rate

Pulse Rate displays the patient's pulse rate as measured with the SpO
5.3.16

I:E

I:E displays the ratio between the length of the inspiration and the length of the expiration.
5.3.17

PScalc

PScalc displays the pressure above the PEEP pressure applied to the patient in PCV/PSV
mode.
5.3.18

Insp. Time

Insp. Time displays the duration of the inspiratory cycle, measured from the start of inspira-
tion to the start of expiration.
5.3.19

Rise Time

Rise Time displays the duration of the pressure or volume increase, measured from the start
of inspiration until the set pressure or volume has been reached.
Doc. 006150 M-1
displays the patient's oxygen saturation, if measured with the SpO
2
sensor accessory, the ventilator displays func-
2
How to Use the Vivo 65
module
2
Sensor:
2
module.
2
User manual