Humidity Comp. Lost; Alarm); Led Failure Alarm; Low Alarm Battery Alarm - Breas VIVO 65 User Manual

Hide thumbs Also See for VIVO 65:
Table of Contents

Advertisement

6.4.21
Humidity Comp. Lost (Humidity Compensation Lost Alarm)
Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
6.4.22

LED Failure Alarm

Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
6.4.23

Low Alarm Battery Alarm

Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
Doc. 006150 M-1
Description

Humidity Comp. Lost

An Humidity Compensation Lost alarm will be given when the
automatic humidity compensation is out of order.
50% relative humidity is used for temporary compensation. If
the ventilator is used at other humidities, delivered and meas-
ured pressure and flow may deviate.
Medium
The ventilator will continue treatment with the same settings.
The accuracy of the volume measurement may be impaired.
Air humidity sensor values (RH and temperature) inside valid
range.
Description
LED Failure
A LED Failure alarm will be given when one or more LED
indicators on the front panel are broken.
Medium
The ventilator will continue treatment with the same settings.
Power-on reset of ventilator (or repair).
Description
Low Alarm Battery
An alarm for Low Alarm Battery will be given if the alarm battery
is not charged enough to have power for a Power Fail alarm for
at least 2 minutes.
Medium
The ventilator will continue treatment with the same settings
and start charging the alarm batteries.
When alarm energy storage level is sufficient to give an alarm
for at least 2 minutes.
Alarms
121
User manual