Spo Sensor Failure / Disconnected Alarm; Spo 2 Artifact - Breas Vivo 3 User Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

7.4.3
SpO
2
This alarm requires a connected SpO
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
7.4.4
SpO
2
This alarm requires a connected SpO
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Doc. 007231 F-3
Sensor Failure / Disconnected Alarm
Description
SpO2 Disconnected
Medium
An SpO
one of the conditions below appears:
Faulty or disconnected sensor.
Normal communication with the sensor is re-established.
An information message remains until acknowledged by the
user.
The Vivo 3 will continue treatment according to the current
settings.
Artifact
Description
Poor SpO2 Signal
Medium
A poor SpO
or artifacts are detected by the sensor.
Check the sensor and its placement on the patient.
An OK signal is received from the sensor or the sensor is
disconnected.
An information message remains until acknowledged by the
user.
The Vivo 3 will continue treatment according to the current
settings.
sensor.
2
Sensor Failure/Disconnection alarm will be given if
2
An error signal is received from the sensor
No signal at all from the sensor is received within 2 seconds.
sensor.
2
Signal alarm will be given if perfusion is too low
2
Alarms 65
User Manual