7.3.12
High SpO
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
This alarm requires a connected SpO
7.3.13
Low SpO
Property
Alarm text
Priority
Definition
Possible cause
Ventilator action
This alarm requires a connected SpO
7.3.14
High Pulse Rate Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
This alarm requires a connected SpO
62 Alarms
User Manual
Alarm
2
Description
High SpO2
Medium
A High SpO
exceeds the alarm limit for 30 seconds.
Too high flow of bleed-in oxygen.
The SpO
The Vivo 3 will continue treatment according to the current
settings.
Alarm
2
Description
Low SpO2
High
A Low SpO
below the alarm limit for 30 seconds.
•
Too low flow of bleed-in oxygen.
•
Oxygen inlet is disconnected.
•
Delivered tidal volumes are too small.
The ventilator will continue treatment with the same settings.
Description
High Pulse Rate
Medium
A High Pulse Rate alarm will be given when the measured pulse
rate exceeds the alarm limit for 15 seconds.
•
Insufficient ventilatory support.
•
Too low flow of bleed-in oxygen.
•
The EPAP value is set too high.
•
Bad positioning of the finger probe.
The pulse rate goes back below the alarm limit.
The Vivo 3 will continue treatment according to the current
settings.
alarm will be given when the measured SpO
2
value goes back below the alarm limit.
2
sensor.
2
alarm will be given when the measured SpO
2
sensor.
2
sensor.
2
2
is
2
Doc. 007231 F-3