7.3.14
High Pulse Rate Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
This alarm requires a connected SpO
Doc. 007229 D-3
Description
High Pulse Rate
Medium
A High Pulse Rate alarm will be given when the measured
pulse rate exceeds the alarm limit for 15 seconds.
•
Insufficient ventilatory support.
•
Too low flow of bleed-in oxygen.
•
The EPAP value is set too high.
•
Bad positioning of the finger probe.
The pulse rate goes back below the alarm limit.
The Vivo 2 will continue treatment according to the current
settings.
30 to 230 bpm (beats per minute)
Off
5 bpm (beats per minute)
sensor.
2
Alarms
103
User manual