6.3.24
Low EtCO
Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
This alarm requires a connected EtCO
6.3.25
High InspCO
Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
110
Alarms
User manual
Alarm
2
Description
Low EtCO2
Medium
A Low EtCO
below the alarm limit for 30 seconds.
•
Alarm limit is set too high.
•
Ventilator disconnection.
•
Excessive leakage in the Patient circuit/Interface.
•
Partial obstruction of the airways.
•
Breath Rate too high.
•
Delivered Tidal Volume too high.
•
Self triggering of the ventilator.
1 to 99 mmHg
Off
1 mmHg
The ventilator will continue treatment with the same settings.
Alarm (High Inspired CO
2
Description
High InspCO2
High
A High Inspired CO
inspired CO
•
Alarm limit is set too low.
•
Excessive dead space between patient and exhalation valve/
leakage port.
•
Leakage port/valve occluded.
1 to 99 mmHg
Off
1 mmHg
The ventilator will continue treatment with the same settings.
alarm will be given when the measured EtCO
2
sensor
2
)
2
alarm will be given when the measured
2
exceeds the alarm limit for 30 seconds.
2
is
2
Doc. 006150 M-1