Measurement Procedure With Autosampler; Clean The Frequency Oscillator For The Next Sample - kruss DS7000 Series User Manual

Density meter
Table of Contents

Advertisement

8.5

Measurement procedure with Autosampler

8.6

Clean the frequency oscillator for the next sample

For automated sample presentation, the sample is transported into the frequency
oscillator and measured after the MEASURE KEy has been pressed and the sample series
entered.
The sample is transported into the measurement system by means of the peristaltic pump.
If the value is stable, it is saved. Once it is saved, the measurement system is rinsed
automatically.
To take the following measurement the rotor is moved to the next sampling position.
After the measurement is taken, the frequency oscillator must be cleaned for the next
sample.
Important note!
Insufficient cleaning of the frequency oscillator leads to sample carryover and
therefore incorrect results.
TIP!
Create a method with a cleaning procedure for this rinsing process. Access this
method and start cleaning by pressing SySTEM > Tab Functions > Function
CLEANING
Cleaning is described in detail in the chapter Initial start-up.
67
Routine operation