Installation
6.8

Connect the density meter to the LIMS

6.9

Connect the density meter to the mains.

your measurement results can be directly transmitted to the LIMS.
1. Connect the density meter with the LIMS using the network cable (1).
2. Configure the network connection in the menuSySTEM > Network.
Attention
Damage to electrical parts due to the build-up of condensation.
h Wait at least one hour after transporting the device from a cool area (e.g.
refrigerated room or outside) before connecting the device to the power supply.
1. Plug the network cable into the density meter (1).
2. Connect the density meter to the power supply using the mains cable.
Permitted voltage source: 90 V...240V, 50/60 Hz
44