10.5

Service tab

Permission
1.1.1

Access service menu

1.1.1

Service explanatory notes

In the Service tab, the following functions can be performed:
• Tare
• Measurement calibration
• Firmware update
• Run SQL script
• Remote monitoring
Device settings and functions may only be modified by users with administrator
permissions.
1. Press SySTEM > Tab Service > EDIT
The service menu opens in edit mode:
Parameters
Tare
156
Explanatory notes
For the calculation of an offset (tare) which is then
automatically included in the result calculation. For this
purpose, the target value (g/cm³) of the present sample
is entered together with the number of measurements
with which the sample is to be measured. A mean value
is calculated from the measurements and the difference
between the measured result and the target value is
calculated. This difference is then automatically taken
into account in all future sample measurements.