Methods menu
3. Enter for Time30 s, Rotation anticlockwise, rpm 50.
4. Activate Auto-confirmation, in order for drying to occur automatically after rinsing.
5. Confirm the entry by pressing OK.
Once it has been rinsed, the frequency oscillator should be dried and then the next step
Step 4: Dry
(test measurement) should begin automatically.
1. Press ADD.
2. Enter 20 s for Time.
3. Activate Auto-confirmation to begin the next step (test measurement) automatically
once drying is complete.
4. Confirm the entry by pressing OK.
After drying, a test measurement should be carried out and it should be dried again if
Step 5: Define test
measurement
there is an incorrect result.
1. Press ADD.
2. Under Module type, select the module Test measurement.
127