Delete Tare; Measurement Calibration - kruss DS7000 Series User Manual

Density meter
Table of Contents

Advertisement

Result
7.1.1

Delete tare

This is the procedure
Result
5.1.1

Measurement calibration

General information
6. Press OK to save the value permanently.
A previously saved tare value will be overwritten.
7. Press CANCEL if the result should not be saved.
The previously saved blank value is then retained.
This tare value is taken into account in the calculation for all measurement results.
An existing tare value can be deleted.
1. In the main menu, press TARE.
or
2. Press SySTEM > TabService > EDIT> TARE.
3. Click on DELETE.
The tare value will be set to zero.
4. Press SAVE.
5. Press OK to delete the value permanently.
The tare value is set to zero.
A measurement calibration is carried out in order to enable correct measurements and to
remove systemic measurement deviations.
For measurement calibration, two measurements (air and water) are taken, in order to
calculate the deviation.
Note!
The frequency oscillator must be very accurately tempered in order to achieve a
precise measurement.
For temperature control, the currently measured temperature is constantly displayed
during tempering.
The temperature display differs as follows:
159
System menu