Methods menu
The cleaning procedure should begin by sucking in the cleaning solution. The next step
should then be carried out automatically.
Step 1: Peristaltic pump
1. Press ADD.
2. Under Module type, select the module Peristaltic pump.
connection
3. Enter for Time30 s, Rotation anticlockwise, rpm 50.
4. Activate Auto-confirmation.
5. Confirm the entry by pressing OK.
Step 2: Pause
The cleaning solution should soak in for 10 s. The hose must then be removed from the
cleaning solution, i.e. auto-confirmation must be deactivated.
1. Press ADD.
2. Under Module type, select the module Pause.