kruss DS7000 Series User Manual page 87

Density meter
Table of Contents

Advertisement

Scale 1 and Scale 2
Scale
Unit
Resolution
The following parameters are defined under METHODS > Scale 1/2 > Scale >-Manage scales- :
Parameters
Default
Entry
Selection menu of pre-defined (marked with *) and user-defined measurement scales.
Density*
Relative density 20/20*
Relative density 20/4*
Alcohol concentration (OIML)*
Brix-Scale*
Sulphuric acid concentration*
- Manage scales-
Units for the pre-defined scales:
• Density > g/cm3
• Relative density for 20/20 and 4/20 > none°
• Alcohol concentration (OIML) > Vol.-% OIML
• Brix > %Brix
• Sulphuric acid concentration > weight-%
Resolution selection list:
Resolutions Scale 1: 0.1; 0.01; 0.001; 0.0001
Resolutions Scale 2: 0.1, 0.01; 0.001; 0.0001
Displays whether a conversion factor is being used.
The value for the conversion factor can once be pre-set as a fixed parameter under
Default or each time directly before measurement (Entry > Yes).
Fixed conversion factor value.
yes/no display,
Yes: a value can be entered prior to every measurement.
No: A fixed assigned value is used to calculate the results.
87
The value actually measured in g/cm3.
Has no unit. The relative density 20/20 is calculated by
dividing the density of the sample at a temperature of
20 °C by the density of pure water at a temperature of
20 °C.
Has no unit. The relative density 20/4 is calculated by
dividing the density of the sample at a temperature of
20 °C by the density of pure water at a temperature of
4 °C.
Change and define user-defined scales.