Methods menu
Explanatory notes
9.7.4

Define intermediate cleaning for autosampler

Example
This is the procedure
This is the procedure
Parameters/keys
The following two parameters (Sampler mode and Interim cleaning procedure) define
the procedure after sample measurement.
Sampler mode
Intermediate cleaning
procedure
Template
If the frequency oscillator is to be rinsed and dried after every measurement, this
intermediate cleaning must be defined.
The cleaning procedure intermediate cleaningis defined in the factory settings.
How to define intermediate cleaning is described in the following example.
The intermediate cleaning of AS80 should include the following steps:
• Step 1: Sampler: rinse, to draw cleaning fluid into the frequency oscillator: Time 30 s,
anticlockwise rotation, rpm 50, continues automatically
• Step 2: Pause, to soften the cleaning solution in the frequency oscillator: 10 s,
continues automatically
• Step 3: Sampler: to rinse the frequency oscillator with water and empty it: Time 30 s,
anticlockwise rotation, rpm 50, continues automatically
• Step 4: Dry to dry the frequency oscillator: 20 s, continues automatically
1. Press METHODS > Tab Sampl. > EDIT.
112
Explanatory notes
Selection menu:
• Displacement: The sample is sucked in and pumped off
after measurement. The next sample then replaces the old
sample. There is no intermediate cleaning.
• Intermediate cleaning: The sample is sucked in and
pumped off after measurement. The hoses are then rinsed
and dried. This intermediate cleaning procedure must be
defined.
Intermediate cleaning between individual samples on
selection of Sampler mode > Intermediate cleaning.
Selection menu:
• none > only for sampler mode Displacement
• already defined cleaning procedures. There are pre-
defined factory-set interim cleaning procedures.
• -Manage cleaning procedure- to define and change
Intermediate cleaning procedures.
Selection menu:
• none
• pre-defined templates
• -Manage templates-
The settings are confirmed, the method menu is closed and
the start menu is opened.
The method menu closes without saving the entry and the
start menu opens.
Deletes the current method.