Methods menu
9.10.2
Select final cleaning
9.10.3
Start final cleaning
9.10.4
TIP for intermediate cleaning
With factory setting values, complete filling of the frequency oscillator is usually
guaranteed. The pumping time and revolution speed should only be modified where
required (e.g. for very viscous samples).
Once a measurement series is complete or before decommissioning, a final cleaning
should be performed.
Where required, this cleaning procedure can be modified and/or newly created.
1. Press METHOD > Tab Cleaning > EDIT.
2. Select final cleaning from Cleaning procedure .
3. Save the selection by pressing OK.
Final cleaning is begun as follows:
1. Access the method with the final cleaning selected under tab Cleaning.
2. Start the final cleaning under SERVICE > Tab Functions > Function Cleaning.
The final cleaning selected in the method will be performed.
If you would like to perform intermediate cleaning between measurements, we
recommend that you define a method that has an intermediate cleaning procedure as
final cleaning.
136