Methods menu
Step 1:
Sampler: Rinse
Step 2:
Define pause
The cleaning procedure should begin by sucking in the cleaning solution. The next step
should then be carried out automatically.
1. Under Module type, select the module Sampler: Rinse.
2. Enter for Time30 s, Rotation anticlockwise, rpm 50.
3. Activate Auto-confirmation to begin the next step automatically.
4. Confirm the entry by pressing OK.
The cleaning solution should soak in for 10 s. The next cleaning procedure should then
begin automatically.
1. Press ADD.
2. Under Module type, select the module Pause.
3. For the pause, enter Time 10 s and activateAuto-confirmation.
114