kruss DS7000 Series User Manual page 160

Density meter
Table of Contents

Advertisement

System menu
This is the procedure
Temperature display
Red
The measured temperature lies above the target temperature.
Blue
The measured temperature lies below the target temperature.
Measured temperature = target temperature.
1. Press SySTEM > Tab Service > EDIT >Measurement calibration.
The dialog box for measurement calibration opens:
Keys
Air / water
160
Explanatory notes
Option Air or Water. Results for air and water must be
present for the calculation
Starts measurement calibration.
Opens a window with the previous calibration results.
Under TABLE the results can be selected and printed in
a sorted manner.
Close the window by pressing OK .
The values are set to Gradient = 1,00000000 and
Offset = 0,00000000.
By pressing SAVE the previous values are deleted and
the above settings are applied.