Define cleaning procedure for the autosampler

Example
This is the procedure
The final cleaning of the autosampler should include the following steps:
• Step 1: Sampler: Rinse, to draw cleaning fluid into the flow cell: Time 30 s,
anticlockwise rotation, rpm 50, automatically continues
• Step 2: Pause, to soak the cleaning solution in the flow cell: 10 s, automatically
continues
• Step 3: Sampler: Rinse, to rinse the flow cell with water and empty it: Time 30 s,
anticlockwise rotation, rpm 50, automatically continues
• Step 4: Dry, to dry the flow cell: 20 s, automatically continues
• Step 5: Test measurement, measurement of empty flow cell. The measurement result
is examined. If the result is not in the defined range, it should be dried and measured
again.
1. Press METHOD > Tab Cleaning > EDIT.
2. Select -Manage cleaning procedure- and press EDIT.