System menu
Upon completion
Result
10.6.1

Run SQL script

Function objective
This is the procedure
Select script
7. Select firmware updates.
8. Begin the firmware update by pressing START.
The firmware update will be performed.
Do not switch off device!
During the update, the device has no functional program to restart.
h Do NOT turn off the device during the firmware update!
9. Wait until the process is complete.
Once the update is complete, a result report will be displayed.
1. Press OK.
2. Remove the USB stick.
3. Restart the device.
The device is now ready with the new device software for further measurements. The
software version is displayed via SySTEM > Service> Firmware or in the main menu
under ? .
Running an SQL script can modify the device's data storage e.g. new methods can be
added or existing ones can be corrected.
Only run scripts if you are sure that they come from the manufacturer.
Before running an SQL script, please create a directory on a USB stick called "SQL" and
copy the script to there.
1. Press SySTEM > Tab Service > RUN SQL SCRIPT.
The SQL Script dialog box opens.
166