1. Fill water into the frequency oscillator.
Measure water
Important!
The water must be free of air bubbles to ensure a correct measurement.
2. Click Measurement calibration
CALIBRATION POINT WATER MISSING is displayed on the button, for as long no
water value is present.
3. Select Water and press NExT.
4. Enter the water target temperature.
5. Press NExT.
If the measured temperature has reached the target value (green) and the temperature is
stable:
6. Press NExT.
The measurement begins.
After the measurement is complete the window with the calculated Offset and Gradient
values is opened.
163