Method: Select Final Cleaning; Start Final Cleaning - kruss DS7000 Series User Manual

Density meter
Table of Contents

Advertisement

Methods menu
9.11.4

Method: Select final cleaning

9.11.5

Start final cleaning

1. Access the required method via METHOD.
2. Select under METHOD > Tab Cleaning > EDIT.
3. Select the required final cleaning under Cleaning procedure .
There are pre-defined factory-set cleaning procedures for final cleaning.
Final cleaning is begun as follows:
1. Access the method with the final cleaning selected under tab Cleaning.
2. Start the final cleaning under SERVICE > Tab Functions > Function Cleaning.
The final cleaning selected in the method will be performed.