kruss DS7000 Series User Manual page 110

Density meter
Table of Contents

Advertisement

Methods menu
This is the procedure
• Direction of rotation clockwise
The pumping time and revolution speed should only be modified where required (e.g. for
very viscous samples).
The pre-defined factory settings are found under:
• METHODS > Method x > Tab Sampl. for sample feeding.
Where required, these settings can be modified.
The peristaltic pump is defined as follows:
1. Click on METHODS > Tab Sampl. > EDIT.
2. Under sample feeding DS7070, select:
3. Enter the Pumping time and Revolution speed where required.
4. Save the entry by pressing OK.
Parameters/keys
Explanatory notes
Pumping time [s]
Pumping time for the sample to be pumped into the frequency
oscillator.
0 to 65535 s, default 10 s
Revolution speed
Revolution speed while the sample is being pumped into the
[1/m]
frequency oscillator.
1 to 50 rpm, default 50 rpm
Direction of
Direction of rotation of the pump for sample uptake: left
rotation
110
Luftschlauch
DS7050
Ablaufschlauch
Abfallschlauch
Einfüllschlauch
Probe
Abfall