Density meter
Table of Contents

Advertisement

Methods menu
3. Enter for Time30 s, Rotation anticlockwise, rpm 50.
4. Activate Auto-confirmation to begin the next step automatically.
5. Confirm the entry by pressing OK.
The cleaning solution should soak in for 10 s. The next cleaning procedure should then
Step 2: Pause
be carried out automatically.
1. Press ADD.
2. Under Module type, select the module Pause.
3. For the pause, enter Time 10 s and activateAuto-confirmation.
4. Confirm the entry by pressing OK.
After the pause, the cleaning solution should be sucked up and the frequency oscillator
Step 3:
be rinsed out with water. The next cleaning procedure (Drying)should then be carried out
Sampler: Rinse
automatically.
1. Press ADD.
2. Under Module type, select the module Sampler: Rinse .