Manual Sample Feeding; Define Sample Feeding With Ds7070 - kruss DS7000 Series User Manual

Density meter
Table of Contents

Advertisement

Explanatory notes
9.7.1

Manual sample feeding

9.7.2

Define sample feeding with DS7070

Sample feeding
Explanatory notes
manual
Manual sample feeding with Luer syringe.
DS7070
Peristaltic pump for semi-automatic filling and cleaning.
Autosampler
Autosampler for fully automatic filling and cleaning.
Our autosamplers AS80 and AS90 are supported.
Keys
The settings are accepted, the method menu is closed and the
start menu opens.
The method menu closes without saving the entry and the start
menu opens.
Deletes the current method.
Creates a new method
For devices without peristaltic pumps, the sample is filled manually into the frequency
oscillator.
Following measurement, the sample is also rinsed off with water or a solvent compatible
with the sample.
Drying is then manually triggered via the function SySTEM > Tab Functions > AIR PUMP .
During sample uptake, the peristaltic pump DS7070 transports the presented sample into
the frequency oscillator. Various throughput speeds are adjustable. Highly viscous liquids
should be transported more slowly to avoid air bubbles being transported along with
them.
The factory setting values for the pump are as follows:
• Revolution speed: 50 rpm
• Pumping time: 10 s