Access function menu

This is the procedure
10.3.2

Available functions

1. Press SySTEM > Tab Functions > EDIT.
The available functions may be found in selection mode.
Available functions
User: Log off
Tare
Peristaltic pump
Air pump
Cleaning
Print: Result
148
Explanatory notes
Logs the user off.
For the calculation of an offset (tare) which is then
automatically included in the result calculation.
For this purpose, the target value (g/cm³) of the
present sample is entered together with the number of
measurements with which the sample is to be measured.
A mean value is calculated from the measurements and
the difference between the measured result and the target
value is calculated. This difference is then automatically
taken into account in all future sample measurements.
The following actions may be performed with the peristaltic
pump:
• Stop
• Start
• Start left
• Start right
• Start Stop
The following actions may be performed with the dry
pump:
• Stop
• Start
• Start Stop
Performs the cleaning procedure that is selected in the
current method.
The function triggers printing.
This is not required if printing occurs automatically.