Initial start-up
7.3.1

Clean frequency oscillator with individual functions

See the chapter System menu for how to assign functions to the keys.
Rinse
1. Fill a rinsing vessel with the rinsing solution or distilled water and hang the filling hose
in the rinsing vessel.
2. Place the waste vessel underneath the waste hose.
3. Press SySTEM > Tab Functions > Peristaltic pump START to transport the liquid into
the frequency oscillator.
4. Rinse the sample chamber with a minimum of 10 ml distilled Water.
5. Stop the pump by pressing SySTEM > Tab Functions > Peristaltic pump STOP if no
more liquid should be sucked in.
Drain
1. Remove the hose from the rinsing vessel.
2. Press SySTEM > Tab Functions > Peristaltic pump START to remove liquid from the
frequency oscillator.
3. Press SySTEM > Tab Functions > Peristaltic pump STOP if no more liquid is dripping
from the waste hose.
air tube
DS7050
drain tube
filling tube
rinsing liquid
air tube
DS7050
drain tube
filling tube
rinsing liquid
48
waste tube
waste
waste tube
waste