kruss DS7000 Series User Manual page 46

Density meter
Table of Contents

Advertisement

Initial start-up
Dry frequency oscillator
To dry the frequency oscillator, prepare the device as follows:
1. Remove the Luer syringe.
2. Connect the drying unit's air hose to the infeed piece on the frequency oscillator.
3. Place an empty waste vessel underneath the waste hose.
Caution
If the waste hose juts into the waste liquid during the drying process, there is a
strong spray.
h Use an empty waste vessel where
h Take care to ensure that the end of the hose does not fall into the liquid.
h Wear suitable protective equipment when using corrosive liquids.
4. Start the drying unit by pressing SySTEM > Tab Functions > Air pump START.
46
air
DS7050
air tube
waste tube
waste
possible.