Initial start-up
7.3.2

Clean frequency oscillator with cleaning procedure

See the chapter Method menu > Tab Cleaning for how to define cleaning procedures.
Rinse
1. Access the method with the defined cleaning procedure steps.
2. Fill a rinsing vessel with distilled water or a suitable solvent and hang the filling hose in
the rinsing vessel.
3. Place the waste vessel underneath the waste hose.
4. Start step 1 of the cleaning procedure with SySTEM > Tab Functions > Function
Cleaning.
The Cleaning function performs the cleaning procedure which is defined in the
Methods.
In step 1, the rinsing solution is sucked up by the peristaltic pump and the peristaltic pump
is stopped when the pumping time is over.
Before the next, second step (sucking up the rinsing solution), the filling hose must be
Drain
removed from the rinsing vessel.
1. Remove the filling hose from the rinsing vessel.
air tube
DS7050
drain tube
filling tube
sample
air tube
DS7050
drain tube
filling tube
rinsing liquid
50
waste tube
waste
waste tube
waste