Measure Quality Controls - kruss DS7000 Series User Manual

Density meter
Table of Contents

Advertisement

Methods menu
Explanatory notes
9.9.1

Measure quality controls

Objective
Define QC method
Measurement
Explanatory notes
options
Interval
measurement can
are begun with the Measure key.
pause
Lock method
The active method can be disabled.
used for measurements.
Keys
The settings are confirmed, the method menu is closed and the
start menu is opened.
The method menu closes without saving the entry and the start
menu opens.
Deletes the current method.
Creates a new method
In order to be able to specially label and filter measured QC samples in the results list,
the QC method must be defined and the measurement option Quality control must be
activated.
1. Access the method for which you would like to measure QC samples.
2. Press EDIT > NEW to create a new method.
3. Enter the name, e.g. QC method 1.
4. Activate Import data of the current method.
All parameters from Method 1 are confirmed.
5. Press OK to save.