Sample uptake and the intermediate cleaning procedure must be defined for the
autosampler specific to method.
There are pre-defined factory settings. Where required, these can be copied and/or
modified.
The sample feeding is defined as follows:
1. Click on METHODS and select the method required.
2. Press tab Sampl. > EDIT.
3. Under Sample feeding Autosampler, select:
Parameters/keys
The following three parameters (pumping time, rotation speedand rotational
direction) are only used for sample uptake into the frequency oscillator.
Pumping time [s]
Revolution speed
[1/m]
Direction of rotation
111
Explanatory notes
Pumping time for sample uptake:
0 to 65535 s, default 10 s
Revolution speed of pump during sample uptake:
1 to 50 rpm, default 50 rpm
Direction of rotation of the pump:
• right > for sample uptake and pumping the sample out
• left > only for rinsing and drying