9.8.2

Define cleaning procedure for DS7070

Example
Example
This is the procedure
How to define a cleaning procedure is described in the following example.
Final cleaning with the peristaltic pump DS7070 is composed of the following steps and
defined:
• Step 1: Peristaltic pump, to draw water into the frequency oscillator:
time 30 s, anticlockwise rotation, rpm 50, continues automatically
• Step 2: Pause, to soften the water in the frequency oscillator 10 s, continues
automatically
• Step 3: Peristaltic pump, to remove water from the frequency oscillator: Time 30 s,
anticlockwise rotation, 50 rpm, Stop once performed (hose must be reconnected
when using drying unit DS7050) or automatically continue once performed (when
using drying unit DS7060, no reconnection is required).
• Step 4: Dry, to dry the flow cell: 20 s, stop once performed
1. Press METHOD > Tab Cleaning > EDIT.
2. Select -Manage cleaning procedure-
3. Press EDIT.
119