Methods menu
Step 3: Peristaltic pump
3. Enter Time10s for the pause.
4. Deactivate Auto-confirmation, in order to remove the hose from the rinsing vessel
before pumping off the cleaning solution.
5. Confirm the entry by pressing OK.
The liquid should be pumped off after the pause. Auto-confirmation is then activated or
deactivated depending on whether the drying unit DS7050 or DS7060 is in use.
• Activated: Stop once performed, hose must be reconnected for use with drying unit
DS7050.
or
• Deactivated: Automatically continue once performed, hoses do not need to be
reconnected for use with drying unit DS7060.
1. Press ADD.
2. Under Module type, select Peristaltic pump.
3. Enter for Time30 s, Rotation anticlockwise, rpm 50.