Install Ds7000 With Drying Unit Ds7050 And Without Peristaltic Pump; Install Ds7000 With Drying Unit Ds7050 And Peristaltic Pump - kruss DS7000 Series User Manual

Density meter
Table of Contents

Advertisement

Installation
6.3

Install DS7000 with drying unit DS7050 and without peristaltic pump

6.4

Install DS7000 with drying unit DS7050 and peristaltic pump

On delivery, the drying unit mentioned in the order, including a cartridge with silica gel
and the infeed and drainage pieces are already installed.
Important note!
If the silica gel is discoloured, it can no longer absorb any moisture and must be
renewed. Replace the silica gel as described in the Maintenance chapter.
All that remains is for the hose connections to be established as follows:
1. Connect the hoses as shown in the overview.
DS7050
filling tube
sample/rinsing liquid
On delivery, the drying unit mentioned in the order, including a cartridge with silica gel
and the infeed and drainage pieces are already installed.
All that remains is for the peristaltic pump and the hoses to be connected.
Important note!
If the silica gel is discoloured, it can no longer absorb any moisture and must be
renewed. Replace the silica gel as described in the Maintenance chapter.
40
air tube
waste tube
waste