Routine operation
8.1.5
Sample numbers
8.1.6
Measure QC sample
8.2
Present sample
8.2.1
Present sample manually
How to define the method is described in the chapter Methods menu
Sample numbers are serially (from 1 to 999) assigned in general.
After sample number 999, the next number is 001.
For the measurement of the QC sample, an individual QC method should be
defined and the measuring option should be activated under METHOD > Tab
Opt. > Quality control.
If the samples to be measured are quality controls:
1. Access the QC method.
The results will be labelled in the result list with QC.
How to define the QC methods is described in the chapter Methods menu > Measure
quality control
Depending on the device type the samples are presented either manually or semi-
automatically.
1. Fill the sample into the frequency oscillator through the filler neck using the Luer
syringe.
The frequency oscillator must be completely filled and free of streaks and gas bubbles.
60