Methods menu
Modules/steps
Explanatory notes
Autosampler: Rinse
For the rinsing of the measurement cell with the autosampler,
the following must be defined:
• Time [s]
• Direction of rotation (clockwise/anticlockwise)
• Revolution speed [1/m]
This module sucks rinsing fluid into the frequency oscillator.
• Auto-confirmation
Autosampler: Rinse
For the rinsing of the autosampler, the following must be
defined:
• Time [s]
• Direction of rotation (clockwise/anticlockwise)
• Revolution speed [1/m]
This module rinses the frequency oscillator with water and
then empties it.
• Auto-confirmation
Dry
For drying, the following must be defined:
• Time [s]
This module switches the air pump on for a defined period of
time.
• Auto-confirmation
Pause
The following must be defined for a pause:
• Time [s]
This module waits for a defined period.
• Auto-confirmation
Test measurement
A test measurement is conducted.
• After test measurement, continue from step:
A test measurement is conducted.
• Auto-confirmation
118
– activated: once rinsing is complete, the next cleaning
module will automatically be performed.
– deactivated: once rinsing is complete, you will be
prompted as to whether the next cleaning module
should be performed.
– activated: once rinsing is complete, the next cleaning
module will automatically be performed.
– deactivated: once rinsing is complete, you will be
prompted as to whether the next cleaning module
should be performed.
– activated: once drying is complete, the next cleaning
module will automatically be performed.
– deactivated: once drying is complete, you will be
prompted as to whether the next cleaning module
should be performed.
– activated: after the pause, the next cleaning module will
automatically be performed.
– deactivated: after the pause, you will be prompted as to
whether the next cleaning module should be performed.
– do not continue
– Selection list: All modules used in the cleaning
procedure are listed. you can choose which step to
continue with.
– activated: once the test measurement is complete, the
next cleaning module will automatically be performed.
– deactivated: once the test measurement is complete,
you will be prompted as to whether the next cleaning
module should be performed.