CLEANING

14 CLEANING

15 SERVICE

102
ProMax X-ray unit with Dimax3
NOTE
When disinfecting the unit surfaces, always
disconnect the unit from mains.
The bite pieces, chin support, chin rest, temple supports
and cephalostat temple supports can be autoclaved up to
135°C or cleaned with alcohol-based solutions.
Other unit surfaces, including the touchscreen, can be
cleaned with a soft cloth damped in a mild cleaning
solution.
To guarantee user and patient safety and to ensure image
quality the unit must be checked and recalibrated by a
qualified PLANMECA service technician once a year or
after every 10 000 exposures if this is sooner. Please
refer to the Planmeca ProMax Technical Manual for
complete servicing information.