Appendix B: Performance Verification
DG2030 Front Panel
B 34
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
CH0
Figure B 13: Delay time measurement connections
Setup.
H Oscilloscope
Displayed channel
Vertical axis
. . . . .
Horizontal axis
. . .
Trigger mode
. . . .
Trigger source
. . . .
Trigger level
. . . . .
Input coupling
. . . .
Input impedance
Characteristics Confirmation Procedure.
1. Load the TP8PULSE.PDA test pattern file from the performance check
disk.
2. Press the START/STOP button on the front panel so that the button LED
indicator turns on.
3. Press the following buttons to set the DG2030 CH0 output delay time to
–5.00 n.
Menu button
Bottom button
SETUP
Timing
Condition
Move the cursor to CH0 with the general purpose knob.
TDS784A Front Panel
Clock Output
BNC Cables
CH1 and CH2
500 mV/div
5 ns/div
Auto
CH1
750 mV
DC
50 W
. .
Pop up menu
Side button
Delay
CH1
CH2
Front panel
button
-5.0
ENTER
DG2030 User Manual