Tektronix DG2030 User Manual page 222

Data generator
Hide thumbs Also See for DG2030:
Table of Contents

Advertisement

DG2030 User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setup.
H Oscilloscope
Displayed channel
Vertical axis
. . . . .
Horizontal axis
. . .
Trigger mode
. . . .
Trigger level
. . . . .
Input coupling
. . . .
Input impedance
H Function generator
Waveform
. . . . . . .
Frequency
. . . . . . .
Amplitude
. . . . . . .
Offset
. . . . . . . . . .
Characteristics Confirmation Procedure.
1. Load the TP2EXCLK.PDA test pattern file from the performance check
disk.
2. Press the START/STOP button on the front panel so that the button's LED
indicator lights.
3. Confirm that the clock pulse waveform can be observed on the oscilloscope.
4. Press the START/STOP button on the front panel to stop output (the
button's LED goes off).
Appendix B: Performance Verification
CH1
500 mV/div
1 ms/div
Auto
500 mV
DC
50 W
. .
Square wave
1 MHz
1 V (50 W termination)
500 mV (50 W termination)