Tektronix DG2030 User Manual page 225

Data generator
Hide thumbs Also See for DG2030:
Table of Contents

Advertisement

Appendix B: Performance Verification
B 14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
BLOCK 2
BLOCK 4
CLOCK OUT
CH0
EVENT OUT
SYNC OUT
CLOCK OUT
CH0
EVENT OUT
SYNC OUT
Figure B 6: Repeat mode timing chart
The following steps check the pattern data output operation in Single run mode.
7. Press the following buttons to set the DG2030 run mode to Single.
Menu button
Bottom button
SETUP
Run Mode
8. Run the oscilloscope so that the oscilloscope is waiting for a trigger event.
9. Press the START/STOP button on the front panel so that the button's LED
indicator lights.
10. Confirm that the CH0 output data pattern as shown in Figure B-7 appears on
the oscilloscope screen once every time the FORCE TRIGGER button on
the DG2030 front panel is pressed.
11. Press the START/STOP button on the front panel to stop output (the
button's LED goes off).
BLOCK 1
BLOCK 3
BLOCK 2
BLOCK 4
Pop up menu
Side button
Single
BLOCK 1
BLOCK 3
Front panel
button