10. Set the inhibit control for the CH2 output channel to OFF as following
procedures:
Menu button
Bottom button
SETUP
Select CH2 with the general purpose knob.
Output
Condition
11. Confirm that the clock output voltage level is almost 0 V.
12. Disconnect the BNC cable from the inhibit input on the DG2030 rear panel.
13. Confirm that the pulses with 10 MHz of frequency, +3 V of high voltage
level and –1 V of low voltage level are continuously output from the
DG2030 clock output.
14. Press the START/STOP button on the front panel to stop output.