Appendix A: Specifications
Table A 2: Warranted environmental characteristics
Characteristics
Temperature
Operating
Non operating
Relative humidity
Operating
Non operating
Altitude
Operating
Non operating
Dynamics
Vibration
Operating
Non operating
Shock
Non operating
Installation requirements
Power consumption
Dissipation (fully loaded)
Surge current
Cooling clearance
Top clearance
Side clearance
Rear clearance
A 6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description
+10 _C to +40 _C
-20 _C to +60 _C
20% to 80% (No condensation)
Maximum wet bulb temperature 29.4 _C
5% to 90% (No condensation)
Maximum wet bulb temperature 40.0 _C
To 4.5 km (15000 ft).
Maximum operating temperature decreases 1_C each 300 m above 1.5 km.
To 15 km (50000 ft).
0.27 Grms, 5 to 500 Hz
2.28 Grms, 5 to 500 Hz
2
294 m/s
(30 G), half sine, 11 ms duration.
Three shocks per axis in each direction (18 shocks total)
300 watts max. Maximum line current is 4 A rms at 50 Hz
90 V libe, with 5 % clipping
x 9 A peak for less than 5 line cycles, at 25 _C, after product has been off for at least 30 s.
2.5 cm
(1 inch)
15 cm
(6 inches)
7.5 cm
(3 inches)