Tektronix DG2030 User Manual page 232

Data generator
Hide thumbs Also See for DG2030:
Table of Contents

Advertisement

DG2030 User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
31. Press the START/STOP button on the front panel so that the button's LED
indicator lights.
32. Confirm that there is a 1 ms trigger event every 10 seconds on the oscillo-
scope channel 2 display.
33. Press the following buttons to set the DG2030 trigger interval to 10 ms.
Menu button
Bottom button
SETUP
Trigger
34. Press the START/STOP button on the front panel to stop output (the
button's LED goes off).
The following steps check that the DG2030 turns off the external trigger signal.
35. Load the TP5EV52.PDA test pattern file from the performance check disk.
36. Press the following buttons to set the DG2030 clock frequency to 400 MHz.
Menu button
Bottom button
SETUP
Clock
37. Press the following buttons to turn off the DG2030 trigger interval.
Menu button
Bottom button
SETUP
Trigger
38. Press the START/STOP button on the front panel so that the button's LED
indicator lights.
39. Make sure that the message Waiting for Trigger is displayed at the upper
right part of the DG2030 screen and the trigger event no longer occurs.
Appendix B: Performance Verification
Pop up menu
Side button
Interval
State (Set to On)
Time
Pop up menu
Side button
Int FREQ
Pop up menu
Side button
Interval
State (Set to Off)
Front panel
button
10
kHz/ms/mV
Front panel
button
4,0,0,MHz/ms
Front panel
button
B 21

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dg2040

Table of Contents