Nikon DS-L4 Instructions Manual page 78

Microscope camera control unit
Hide thumbs Also See for DS-L4:
Table of Contents

Advertisement

Retracting the Elevating Section and Moving the Stage to the Specimen Removal
1.18.3
Position
Prerequisite for motorized operation
A motorized XY stage must be attached to the microscope in order to perform the following operation.
[Procedure]
(1)
Press the [XYZ] button.
(This operation is not required if the [ESCAPE] button
is already displayed on the [MIC EASY] screen. Go to
the next step.)
(2)
Press the [Specimen Removal Position] button.
The elevating section is moved by the amount specified for
the [ESCAPE DISTANCE]. The motorized XY stage is also
moved to the software limit position at the front of the
microscope.
For details on the escape distance, refer to Chapter 6, "5.2
Setting the Retracting Amount of the Elevating Section".
For details on the software limit, refer to Chapter 6, "6.4
Setting the Software Limits".
The following operations are disabled when moved to
specimen removal position.
-
Focus knob of the microscope, ergo controller, and
joystick
-
XY direction movement with the ergo controller or
joystick
-
[ESCAPE] button
Also, the display of X, Y, Z coordinates on the
[MICROSCOPE CONTROL] screen blinks.
Press the button again to return the motorized XY stage to
the original position and enable XY direction movement
with the ergo controller and joystick.
The elevating section is not returned to the original
position, but escape state is canceled and the focus knob
is enabled.
In addition, the [ESCAPE] button is enabled.
Chapter 5
Individual Operations
[MICROSCOPE CONTROL] Screen
[MIC EASY] Screen
1-66