(5)
Press the button for address 1 to bring the objective in
address 1 into the optical path.
(6)
Use the focus knob on the microscope to focus on the
specimen.
(7)
Press the [SET] button to save the focal point of the
address 1.
The configuration status is shown in the [FOCUS
POSITION] section on the right.
[set]: Focal point has been configured
[not set]: Focal point has not been configured
(8)
Save the focal points for address 2 and later in the
same way.
For addresses where no objective is
attached
Be sure to set the focal points for all nosepiece
addresses. For addresses where no objective is
attached, save the focal point with the same status
as the previous address. If you don't set the focal
point, parfocal correction is not performed when
the address is switched from the position without
objective to the position with objective, or vice
versa.
(9)
When you have saved focal points for all addresses,
press the [OK] button.
(10) Set [AUTO LINK FOCUS] to [ON].
The parfocal correction is turned on.
(11) Press the [OK] button.
[MOVEMENT] screen is displayed.
After making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save the settings to the
microscope's memory. If the microscope is turned off without saving settings, previous values are restored. For more
information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.