Configuring the Connection of Motorized Shutter

3.2
When changing the usage of the motorized shutter to AUX (other than EPI/DIA), configure the motorized shutter
connection with this procedure.
When the intended shutter is disabled
If the dia-illumination shutter opens/closes when the [SHUTTER EPI] button is pressed or epi-illumination shutter
opens/closes when the [SHUTTER DIA] button is pressed, the motorized shutter is not correctly connected to the
EPI SHUTTER/DIA SHUTTER connector.
In this case, turn off the microscope and DS-L4, then reconnect the motorized shutter to the EPI SHUTTER/DIA
SHUTTER connector of the microscope (the control box B for Ni-U). Or, check the usage of the motorized shutter
connected to the EPI SHUTTER/DIA SHUTTER connector and change the usage with this configuration.
Setting Item
EPI SHUTTER PORT
(at the rear of the microscope for
Ni-E, and the control box B for Ni-U)
DIA SHUTTER PORT
(at the connector box for Ni-E, and
the control box B for Ni-U)
[Procedure]
[MIC SETUP] → [CONNECTION]
(1)
Press the button for the connector to be configured.
A sub screen opens to select the function of the shutter.
(2)
Select the function of the shutter ([EPI], [DIA] or
[AUX]).
(3)
Repeat steps (1) and (2) to configure the other
connector.
After making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save the settings to the
microscope's memory. If the microscope is turned off without saving settings, previous values are restored. For more
information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.
Chapter 6
Settings
Setting Value
[EPI] (default)
[DIA]
[AUX]
[EPI]
[DIA] (default)
[AUX]
1-95
Description
For epi-illumination
For dia-illumination
For auxiliary shutter
For epi-illumination
For dia-illumination
For auxiliary shutter