Changing the Digital Camera Operated with the Microscope's CAPTURE Button
4.2.3
The capture method when the CAPTURE button of the microscope is pressed can be changed.
Setting Item
[CAPTURE BUTTON]
[Procedure]
[MIC SETUP] → [BUTTON FUNC]
Choose the capture method among the options in the
[CAPTURE BUTTON].
The displayed buttons depend on the connection
configuration of the digital camera. See Chapter 6, "3.1
Configuring the Connection of Digital Camera".
For Ni-E and Ni-U, after making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save
the settings to the microscope's memory. If the microscope is turned off without saving settings, previous values are
restored. For more information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.
For Ci-E, you cannot save changes to the settings made here. Pressing of the [SAVE] button is not necessary.
Chapter 6
Setting Value
[OFF]
[USB]
(Default for Ni-E/Ni-U)
[FRONT]/ [LEFT]/
[RIGHT]/[REAR]/[AUX]
(DSC connector)
(Default for Ci-E)
Settings
Capture Method
Disables the operation of the CAPTURE button.
Capturing by sending a communication command from the
USB connector
Capturing by output of trigger signal from the DSC connector
(Following screens are for Ni-E)